Type 202755
JUMO CTI-500
Is the JUMO CTI-500 electrically isolated?
Yes, the measuring cell and the analog outputs as well as the voltage supply and the switching outputs are electrically isolated.
What is a CIP process (Clean in Place/ Process)?
CIP is a concept, in which the cleaning and disinfection dissolution circulates in the circuit without dismantling the system to be treated. This is called a circulation, circuit or recirculation cleaning process that is mainly controlled automatically by programs.
The CIP cleaning concept can be used in systems where circuits can be established, e.g. in tank systems, process lines for the drinking milk production, the butter production, the cheese production, the dried milk production or beverage industry (e.g. in breweries).
The task of the conductivity measuring technology is to control the concentration of the cleaning and disinfection solutions in the supply containers and at decisive points during the process. The concentration in the lye and acid container must kept constant by intensification with concentrated lye or acid. The conductivity of the hot water must be controlled because this is the fluid that comes into contact with the product.
What is a desalination function or the biocide addition?
The desalination function is an automatic process, during which the reference value is reduced by opening a desalination valve after the set limit value is exceeded.
The water supply that is now started (level control) dilutes the water in the circuit and, thus, reduces its conductivity.
In cooling towers, for example, biocides are added to the cooling water to prevent the growth of biological substances in the cooling systems. The desalination lock is started after the dosing time to ensure that the biocide does not enter the sewage system.
Which measuring cell and which transmitter of the separated version belong together?
Ensure that the 19-digit F No. on the transmitter rating plate complies with the 19-digit F No. on the measuring cell cable end.
If the numbers do not coincide, calibrate the transmitter and the measuring cell using the calibration set (sales article number 20/00459436) (see point 3 in the FAQs).
What are the repair possibilities?

For the head transmitter:
Repair is only possible in the factory.

With separate cell:
The measuring probe can be replaced. After replacement, calibrate the probe using the calibration set (sales article number 20/00459436) (see point 3 in the FAQs).

What is the correct fitting position for the JUMO CTI-500?
  • The JUMO CTI-500 can be fitted at any position.
  • Avoid direct sun radiation.
  • When installed in a pipe system, ensure a minimum distance of 20 mm between probe and pipe wall to achieve a good flow through  and around the probe.
  • If the minimum distance cannot be kept, the "Fitting factor" parameter allows for limited compensation.
How can the measuring cell for the transmitter be calibrated?
Calibration of the measuring cell is only required after the measuring cell or the transmitter has been replaced.
For the procedure, please refer to the calibration instructions.
How to change the operating language?

The language can be changed with the setup program or directly on the device.
From the device software 196.02.01, eight device languages are provided: German, English, Spanish, French, Polish, Swedish, Italian and Portuguese.
Older devices cannot be updated to all current languages.

Changing the language with the setup program:

  1. Open the setup program
  2. Connect the device to the setup
  3. Read the data from the device
  4. Select the device data
  5. Change the language
  6. Transmit the data back to the device

Changing the language on the device:
The device language is changed in the operator level.

  1. Press the PGM key for more than 3 seconds
  2. Activate the device at the administrator level (standard password 300 can be changed by the customer)
  3. Use the PGM key to select the parameter level
  4. Use the arrow key to go to the device data parameter and select with the PGM key
  5. Use the PGM key to select the language parameter
  6. Select the desired language with the arrow keys and confirm with the PGM key
  7. Press the EXIT key to exit the parameter level
Where should the separated version be used?
In pipe systems subject to vibrations and high temperatures.
teaser
manual
m
https://do2p1q9b92sgp.cloudfront.net/rendition/353604884319/image_p8m78bcjpl6j978npbc5vkkj36/-C2048x1152.6994535519125,0,134.4
Man with smartphone in front of a laptop
Or send us a message
Contact us and we will answer your questions – whether about basics or specific topics.
/content/jumo-website/dk/da/services/faq/faq-form
Send a message now