Other possible connection errors and their causes
- The indicating instrument indicates these mperature of the room
Damaged thermocouple or cable
- Indication consistent as to value, but with a minus sign
Swapped polarity in the pointing device.
- The indicated temperature is clearly too high; drift indication
a) Swapped polarity of the equipotential bonding wire on the connection head (two additional thermocouples are created as a result of swapping wires)
b) Incorrect expansion line
- Indication is clearly too high or too low
a) Incorrect linearisation in the pointing device.
b) Incorrect expansion line
- Indication is too high or too low by a fixed value
Incorrect comparison point temperature
- Indication correct, but slowly drifting
The temperature of the comparison point is unstable or has not been measured.
- The display contains an error of 20 to 25 °C
Linearisation of an L-type thermocouple as a J-type or vice versa.
- When one pole of the thermocouple is disconnected, a value is still indicated
a) Electromagnetic voltages interfere with the input cable
b) Due to the lack of galvanic separation and insufficient insulation, interference voltages have appeared, penetrating e.g. through the furnace insulation.
- Even after disconnecting both poles of the thermocouple, a high value is still indicated
a) Electromagnetic interference interferes with the input cable
b) Galvanic voltages, caused by e.g. damp insulation, interfere with the equalizing cable