pH sensor
product overview
Accurate pH measurement is essential for reliable process control in water treatment, wastewater treatment, food production, pharmaceuticals and chemical processes. The right combination of pH sensor, transmitter and controller contributes to stable product quality, efficient process operation and optimum control over critical processes.
JUMO offers a comprehensive range of pH sensors, pH electrodes, pH controllers, transmitters and accessories for a wide variety of industrial applications.
Thanks to a variety of electrode designs, materials, process connections and communication options, there is always a suitable solution for your installation.
pH sensors
Whatever pH sensor your process requires, JUMO supplies reliable and accurate pH electrodes for continuous pH measurements in industrial applications. Our range includes pH sensors for water and wastewater treatment, chemical processes, food production, pharmaceutical plants and laboratory applications.
Thanks to a variety of membrane types, reference systems, process connections and materials, a suitable pH electrode can be selected for virtually any application. This ensures high measurement accuracy, a long service life and stable process control, even under harsh process conditions.
pH controllers and transmitters
Accurate pH measurement starts with a good sensor, but is only complete with a reliable signal conditioner or controller. JUMO offers an extensive range of pH controllers, analysis transmitters and signal conditioners for visualising, monitoring and automatically controlling pH processes.
Our systems process the sensor signal into a stable measured value, can generate alarms in the event of deviations and, if required, control dosing pumps or control valves. As such, they form an essential part of modern process automation within water treatment, the chemical industry, the food industry, the pharmaceutical sector and utility installations.
Whether you wish to integrate a simple pH measurement into a PLC system or require a complete liquid analysis platform, JUMO offers a suitable solution for every application.
pH accessories
For reliable and long-term accurate pH measurement, the right accessories are just as important as the sensor itself. JUMO supplies an extensive range of accessories for the installation, calibration, maintenance and optimisation of pH measurement systems. These include calibration buffer solutions, temperature sensors for automatic temperature compensation, cables, connectors, protective caps, storage fluids and mounting brackets.
With the right accessories, you can extend the service life of the pH electrode, reduce maintenance costs and increase the reliability of the measurement results. This ensures that your pH measurements remain accurate and reproducible, even under challenging process conditions.
Frequently asked questions
Here you will find answers to the most frequently asked questions about pH sensors
How does a pH sensor or pH electrode work?
What is the difference between a digital and an analogue pH sensor?
What is the importance of a reference elektrode?
How long does a pH sensor last?
How can you tell if a pH elektrode is worn out?
Do you need to calibrate a pH sensor?
How do you calibrate a pH sensor?
How often should a pH sensor be calibrated?
What happens if a pH sensor is not calibrated?
What calibration procedures are used for pH sensors?
Which buffer solutions are best for calibrating a pH sensor?
How do you measure the pH value?
What does the pH value mean?
Why is temperature important when measuring pH?
What is the pH value of water?
Why is the pH value of water important?
What is a good pH value for industrial water?
<br/>
How does a pH sensor or pH electrode work?
A pH sensor, also known as a pH electrode, measures the activity of hydrogen ions in an aqueous solution. This activity determines how acidic or alkaline the solution is. The pH-sensitive glass membrane generates an electrical potential that depends on the hydrogen ion activity. The potential difference between the measuring electrode and the reference electrode is converted into a pH value by a pH meter, transducer or controller.
For most aqueous solutions, the pH scale ranges from 0 to 14. At 25 °C, a pH of 7 is neutral. Values below 7 are acidic, and values above 7 are basic or alkaline.
<br/>
What is the difference between a digital and an analogue pH sensor?
An analogue pH sensor provides a raw millivolt output which is processed by a transmitter or controller. The quality of the measurement can be affected by cable lengths, electromagnetic interference and moisture.
In a digital pH measurement system, the analogue electrode signal is digitised close to or within the sensor. The measured value and, depending on the system, additional sensor and diagnostic data are transmitted digitally to a transmitter or control system. This reduces the impact of electromagnetic interference and long analogue signal cables, and lowers susceptibility to interference. As a result, diagnostic functions and sensor identification become possible, and functions for condition monitoring and predictive maintenance can be supported.
Digital pH sensors are therefore particularly suitable for installations where reliability, availability and ease of maintenance are important.
<br/>
Why is a reference electrode important?
A pH measurement is based on the comparison between a measuring electrode and a reference electrode. The reference electrode provides a stable reference signal which can be used to determine the actual pH value.
If the reference electrode becomes contaminated or damaged, measurement accuracy decreases. A properly functioning reference system is therefore essential for reliable and reproducible pH measurements.
<br/>
How long does a pH sensor last?
The service life of a pH sensor can vary from a few months to several years. The actual service life is determined by factors including process temperature, chemical exposure, pressure, fouling, cleaning procedures, storage and maintenance. In many industrial applications, a service life of approximately one to three years is achieved, but this must always be assessed on a case-by-case basis.
Factors that influence service life include:
• High temperatures
• Extreme pH values
• Chemically aggressive media
• Contamination of the membrane
• Incorrect storage
• Inadequate maintenance
<br/>
How can you tell if a pH electrode is worn out?
A pH electrode that needs replacing often displays one or more of the following characteristics:
• Slow response time
• Difficult or unstable calibration
• Significantly deviating readings
• Reduced measurement accuracy
• Visible contamination or damage to the glass
If an electrode no longer provides stable readings despite cleaning and calibration, replacement is usually the best solution.
<br/>
Do you need to calibrate a pH sensor?
Yes. A pH sensor must be calibrated regularly to ensure measurement accuracy. The measured value may gradually deviate due to electrode ageing, contamination and process influences.
Calibration determines and corrects the zero point and sensitivity of the measuring system. If contamination, damage or ageing affects performance too significantly, the electrode must first be cleaned, conditioned or replaced.
<br/>
How do you calibrate a pH sensor?
In a standard calibration, the pH electrode is immersed in one or more buffer solutions with a known pH value.
The most common procedure is as follows:
1. Clean the electrode.
2. Rinse the sensor with clean water.
3. Place the sensor in a neutral buffer, for example pH 7.00 or pH 6.86, depending on the buffer set used.
4. Carry out the first calibration step.
5. Select the second buffer based on the expected measurement range. For example, use pH 4 and pH 7 for acidic applications, or pH 7 and pH 10 for alkaline applications. For a reliable calibration, the selected buffer values should ideally enclose the expected process pH.
6. Place the sensor in a second buffer.
7. Complete the calibration in accordance with the measuring instrument’s instructions.
For industrial applications, a two-point calibration is usually used.
<br/>
How often should a pH sensor be calibrated?
There is no universal calibration interval that is suitable for every application. The required frequency depends on the process medium, the temperature, the required accuracy, the sensor load and the stability of previous calibration results. For a new or unknown application, start with frequent checks and only extend the interval once recorded calibration data shows that the sensor remains stable.
As a guide, you can use the following guidelines:
• Laboratory applications: daily to weekly
• Water treatment: weekly to monthly
• Critical processes: in accordance with quality or validation procedures
• Stable industrial processes: monthly or at each maintenance interval
Regular checks using buffer solutions help to determine the optimal calibration interval.
<br/>
What happens if a pH sensor is not calibrated?
If a pH sensor is not calibrated regularly, measurement errors may occur. This can lead to incorrect process control, inaccurate chemical dosing and quality issues within the production process.
For critical applications, this can result in higher operating costs, product rejection or unplanned downtime.
<br/>
What calibration procedures are used for pH sensors
Depending on the required measurement accuracy and the application, various calibration procedures may be used.
Single-point calibration: In a single-point calibration, the zero point or offset of the measuring system is primarily corrected. The actual electrode slope is not fully determined in this process. This method is particularly suitable for checks or applications with a limited measuring range. For accurate process measurements, a two-point calibration is generally preferable.
Two-point calibration: A two-point calibration uses two buffer solutions, for example pH 7 and pH 4 or pH 10. This allows both the zero point and the slope of the electrode to be checked. This is the most commonly used calibration method in industrial pH measurements.
Three-point calibration: A three-point calibration uses three buffer solutions, for example around pH 4, 7 and 10. This allows the electrode’s performance to be checked over a wider measuring range. However, for processes with a limited pH range, a correctly performed two-point calibration may be sufficient.
Automatic calibration: Modern digital pH measurement systems can guide users step by step through the calibration process and automatically save the calibration data. This reduces the risk of user error and supports traceability within quality and validation processes.
The most suitable calibration method depends on the application, the required measurement range and the accuracy requirements of the process.
<br/>
Which buffer solutions are used to calibrate a pH sensor?
Buffer solutions with a known pH value are used to calibrate a pH sensor. Commonly used buffer solutions have nominal values around pH 4, pH 7 and pH 9 or 10. The exact buffer value depends on the type of buffer and the temperature. You should therefore use the temperature values provided by the buffer manufacturer and select the correct buffer set in the measuring instrument.
• pH 4 for acidic applications
• pH 7 as a neutral reference point
• pH 10 for basic or alkaline applications
For the best results, it is recommended to choose buffers that are as close as possible to the normal measurement range of the application. It is also important to use fresh and uncontaminated buffer solutions to ensure that the calibration remains accurate and reliable.
<br/>
How do you measure the pH value?
The pH value is measured by bringing a pH electrode into contact with the liquid to be measured. The sensor generates an electrical signal that corresponds to the acidity of the medium.
This signal is then processed by a pH meter, transmitter, controller or process control system, after which the current pH value is displayed.
<br/>
What does the pH value mean?
The pH value indicates how acidic or basic a liquid is.
• pH 7 = neutral
• pH below 7 = acidic
• pH above 7 = basic or alkaline
The pH value plays an important role in chemical reactions, biological processes and the quality of products and process water.
<br/>
Why is temperature important when measuring pH?
Both the sensitivity of the pH electrode and the actual pH of the medium depend on temperature. Automatic temperature compensation corrects for the temperature-dependent electrode slope and takes the temperature of the buffer solution into account during calibration. However, this does not always compensate for the actual change in the pH of the process medium caused by temperature. It is therefore important to record the measurement temperature when carrying out precise measurements.
<br/>
What is the pH of water?
In theory, pure water has a pH of 7 at 25 °C and is considered neutral.
For drinking water, a pH range of approximately 6.5 to 8.5 is often used. However, the applicable requirements and target values vary by country, application and water supplier.
<br/>
Why is the pH of water important?
The pH has a direct impact on corrosion, limescale, chemical reactions, microbiological processes and the operation of water treatment systems.
That is why the pH is continuously monitored and controlled in many industrial processes.
<br/>
What is a suitable pH level for industrial water?
The optimum pH level depends on the application.
Examples include:
• Wastewater: depending on regulatory requirements
• Cooling water: depending on the water treatment programme
• Process water: depending on the process requirements
The desired pH level is always determined by process requirements, regulations and plant conditions.